[Servicebio] Chemiluminescence Imaging System

Cat-No. SCG-W3000

第8页-7.PNG


5. Preview and Capture Pages

In the preview page, you need to enter the location where the experimental results will be stored. The file name is an optional input to facilitate file searching for the user.

The user can choose between manual exposure and automatic exposure as the current experimental mode. For manual exposure, the user needs to input the exposure time for the exposure. For automatic exposure, the algorithm calculates the optimal exposure time.

On the right side of the preview page, there is a time input field in microseconds (us), which represents the exposure time for the bright field image. Clicking the preview switch will start the preview, and the preview time can be adjusted as needed.

Shooting mode:High quality, standard, high sensitivity. You can choose a mode to shoot according to your needs, the high-quality mode takes the best image quality, the standard mode takes into account the image quality and shooting time, and the high-sensitive mode takes the shortest shooting time.

Clicking the capture button will initiate the capture and exposure process, while clicking the back button will return to the main page.

ServiceBio는 생명과학 및 의학 연구에 사용되는 고품질 실험 장비와 시약을 제공하는 글로벌 생명과학 기업입니다.





제품 설명 


Chemiluminescence Imaging System

 



제품 번호


SCG-W3000




제품 특징


Product Introduction


SCG-W3000 is an imaging device that utilizes chemiluminescence technology and is equipped with a high-sensitivity cooled camera with 9 million pixels. It enables fast, accurate, and high-throughput detection and imaging of samples, and is widely used in fields such as life sciences, medicine, and environmental protection.


II. Features

Real-time Imaging: Present the changes in sample signals during the exposure process in real-time, capturing every detail of the shot. Overexpo
sed areas will be indicated for samples that are overexposed.

Time Imaging: After the exposure is completed, every frame image within the exposure time can be generated. Through precise retrospective adjustments, users can select any frame image within that exposure time as the final output.

Time Accumulation: For samples with insufficient exposure, users can choose to continue the exposure after the exposure ends, allowing the sample to obtain additional exposure effects based on the exposure time already accumulated.

Automatic Exposure: Intelligent exposure technology can quickly determine the optimal exposure time. With the functions of time imaging and time accumulation, users can obtain the best image results with just one operation.


SWE Imaging Grayscale Analysis Software-SCG-W2000 SCG-W3000 (W3000 Plus)-3.3.2-win32-x64-setup.exe

Download link from Google Drive:https://drive.google.com/file/d/1pQM6mudvBcxWcDcphP6xECavH_oKfwk_/view?usp=sharing


SWE Imaging Grayscale Analysis Software-SCG-W2000 SCG-W3000.pdf



image.pngimage.png

image.png


image.png


image.png


III. Technical Specifications and Precautions 

Product Name

Chemiluminescence Imaging System

Cat. No.

SCG-W3000

Dimensions

400mm×371mm×700mm

Camera

Pixel Resolution

9 million pixels

Resolution

3000×3000

Pixel Size

3.76μm×3.76μm

Target Size

1 “(11.28mm×11.28mm)

Full Well Capacity

16.5ke-(HCG),50.5ke-(LCG)

Sensitivity

877mv@1/30s

Readout Noise

1.24e-(HCG), 3.22e-(LCG)

Dark Current

0.0003e-/s/pixel@-15℃

Signal-to-Noise Ratio

42.2dB(HCG), 47dB(LCG)

Exposure Time

0.1ms~1h

Binning Mode

1×1, 2×2, 3×3

Grayscale

16-bit (65536 levels)

Cooling

Relative to Ambient Temperature-40℃

Camera Type

Black and White Camera

Lens

Aperture

F0.95-F16

Focal Length

17mm

Type

Fixed Focus Lens

Light Source

Bright Field Light Source

Downward-facing LED white light source, symmetrically distributed on both sides

Dark Box

Light Isolation

Fully light-sealed, isolates environmental light.

Door Control

Door control sensor can control the on/off of the bright field light source.

Field of View

Effective field of view is 136mm*136mm (expandable to 200mm×200mm if needed).

Software Functions

Software Functions

Real-time Imaging: Real-time presentation of changes in sample signals during exposure process, capturing every detail of the capture.

Overexposed areas can be displayed.

After exposure, each frame image within the exposure time can be generated. Users can select any frame image for the final output by precise retrospective adjustment.

Time Imaging

For samples with insufficient exposure, users can choose to continue exposure after the exposure ends, allowing the sample to gain additional exposure effect based on the already exposed time.

Auto Exposure

Intelligent exposure technology quickly determines the optimal exposure time. With time imaging and time accumulation functions, users can obtain the best image result with just one operation.

Industrial Computer

10.4 inches, 1024×768, Windows operating system

External Interfaces

2 USB 3.0 ports

Operating Voltage

100V-240V

Product Power

100W

Product Net Weight

25 Kg


Notes:

1. It is prohibited to touch or scratch the internal lenses of the dark box with hands or sharp objects.

2. After placing the experimental samples, make sure to close the instrument's flip door to prevent external light from entering the dark box and affecting the experimental results.

3. During imaging experiments, shaking the experimental table or instrument is prohibited to avoid impacting the image quality.

4. Pay attention to electrical safety. Pulling or moving the power cord during the experiment is prohibited.

5. After the experiment is completed, clean the samples and any residues inside the dark box thoroughly.

Cat. No.SCG-W5000SCG-W3000SCG-W1000
Dimension400×371×700 mm400×371×700 mm400×371×700 mm
CameraDepth-cooled high sensitivity cameraDepth-cooled high sensitivity cameraHigh-sensitivity camera
Resolution3000×3000,9 megapixels3000×3000, 9 megapixels3072×2048,6.3 megapixels
Pixel3.76×3.76 μm3.76×3.76 μm2.4×2.4 μm
Shooting AreaEffective field of view for blotting film/protein gel: 136×136 mm (can be expanded to 200×200 mm if required);
Effective field of view for 
nucleic acid gel: 200×200 mm.
Blotting Film
136×136 mm (expandable to 200×200 mm if required)
Nucleic Acid Gel / Protein Gel
140×140 mm
Cooling TemperatureRelative ambient temperature -40°CRelative ambient temperature -40°C-
Light SourceBright-field Light Source: Downward-facing LED white light source, symmetrically distributed on both sides.
UV Light Source: 310 nm LED array for uniform transmission illumination.
Downward-facing LED white light, symmetrically distributed on both sidesBright-field Light Source: Downward-facing LED white light source, symmetrically distributed on both sides.
UV Light Source: 310 nm LED array for uniform transmission illumination.
Industrial Computer10.4 inches, 1024×768
Windows operating system
10.4 inches, 1024×768
Windows operating system
10.4 inches, 1024×768
Windows operating system
External Interface2 USB3.02 USB3.02 USB3.0
Working Voltage100 V-240 V100 V-240 V100 V-240 V
Product Power100 W100 W100 W
Net Weight30 kg25 kg30 kg
Real-Time ImagingYesYes-
Time ImagingYesYes-
Time AccumulationYesYes-
Auto ExposureYesYesYes
Choice of 3 Imaging ModesYesYes-
Protein Gel/Nucleic Acid Gel ImagingYes-Yes
Nucleic Acid Gel CuttingYes-Yes


IV. Functional Description




企业微信截图_17089192486253.png

① Internal camera lens assembly, which is the core component of the imaging system.

② 10.4-inch industrial computer with user software operating interface.

③ High-quality light-shielded imaging dark box.

④ Sample tray, removable, used for placing samples to be photographed.

⑤ LED white light source for illumination.

⑥ Power socket and switch.

⑦ Cooling device.

⑧ External USB 3.0 interface.


V. Operating Procedures

1. Power On

Plug in the power cord and turn on the power switch at the back of the instrument. The industrial computer will start up.

2. Sample Loading
Open the instrument door, take out the sample tray, place the prepared text sample on the tray, and then place the tray flat in the groove inside the instrument dark box. Close the instrument door.

3. Launching the Chemiluminescence Imaging Software

After the industrial computer starts up, the chemiluminescence application software will be automatically loaded. Once the software is successfully launched, it will navigate to the main page.

The top-left section displays the company logo.

The bottom-left section is the status bar, showing the current camera and control board connection status, as well as detection of the inserted removable disk.

The top-right section displays the software version number.

The bottom-right section includes buttons for switching between simplified/traditional Chinese, switching to English, exporting the page, and closing the program.

Clicking on the central icon will enter the preview and capture page.

第7页-4.PNG 


4. Software Self-Check
If the software starts abnormally, or if the camera or control board is not connected, the text in the software system will turn gray, and the central icon will also become gray. The self-check completion prompt will turn red, and it will not be possible to click into the preview and exposure page. At this time, you need to check whether the camera and PCB board drivers are correctly installed.


第8页-7.PNG


5. Preview and Capture Pages

In the preview page, you need to enter the location where the experimental results will be stored. The file name is an optional input to facilitate file searching for the user.

The user can choose between manual exposure and automatic exposure as the current experimental mode. For manual exposure, the user needs to input the exposure time for the exposure. For automatic exposure, the algorithm calculates the optimal exposure time.

On the right side of the preview page, there is a time input field in microseconds (us), which represents the exposure time for the bright field image. Clicking the preview switch will start the preview, and the preview time can be adjusted as needed.

Shooting mode:High quality, standard, high sensitivity. You can choose a mode to shoot according to your needs, the high-quality mode takes the best image quality, the standard mode takes into account the image quality and shooting time, and the high-sensitive mode takes the shortest shooting time.

Clicking the capture button will initiate the capture and exposure process, while clicking the back button will return to the main page.



第8页-6.PNG


Preview and Capture Pages


6. Capturing Process

(1) Select automatic exposure and set preview time.

Automatic exposure: Intelligent exposure technology can quickly determine the optimal exposure time.

(2) Real-time imaging.

Click the capture button to start the exposure. The left side displays the stripe, while the right side shows the countdown of the exposure time. As the countdown progresses, the stripe on the left side will show real-time imaging.

Real-time imaging: Real-time display of changes in sample signal during the exposure process, enabling users to grasp every detail of the capture process. This breakthrough function not only improves capture efficiency but also greatly enhances user interaction experience.

During the real-time imaging process, any overexposed areas in the stripe will be displayed in red. If the stripe is judged to meet requirements, the user can click the stop button in the lower right corner to end the exposure early.

第9页-10.PNG


Real-time imaging



(3) Temporal imaging, temporal accumulation.

After the exposure is completed, it automatically enters the results page where adjustments can be made to the captured results.

Time imaging:Through precise retrospective adjustments, users can select any frame within the exposure time as the final output result.

Time accumulation:Even after the exposure is completed, users can choose to continue exposure to obtain additional exposure effects based on the already exposed time. When clicking on "continue exposure," the user will be prompted with the minimum exposure time. The set time needs to be greater than this minimum time. If the set time is shorter than this minimum time, the actual exposure time will be the suggested minimum exposure time.



第10页-11.PNG 


Figure 4.5 Temporal imaging

第10页-12.PNG 


Temporal accumulation


(4) Image adjustments, result saving. After obtaining satisfactory images through automatic exposure, continued exposure, and temporal imaging, adjustments can be made to the bright field image, dark field image, and composite image on the results page.

① Contrast:Adjust the brightness and darkness of the bright field image.

② Exposure:Adjust the lightness and darkness of the dark field image. Clicking on "auto-adapt" can recommend the most suitable value.

③ Rotation: Rotate the image.

④ Invert:Apply an inverted color effect to the image, turning black into white and white into black.

⑤ Crop: Clicking on crop will generate a region on the image. This region can be resized, and after selecting the appropriate size, clicking on crop again will crop out the desired area.

⑥ Vertical flip:Flips the image vertically.

⑦ Horizontal flip:Flips the image horizontally.

⑧ Restore:Restores the image to its original state.

⑨ Import: Select the file automatically saved by the program to import it into the program for further operations.

⑩ Image saving: Save the image and also save the bright field image, dark field image, and composite image.

⑪ Result saving: Save the adjusted image as an encrypted file. 

If any adjustments are made to the image on the results page


第11页-14.PNG 


 Bright field image


第11页-13.PNG 


Dark field image

 

第12页-16.PNG 

Composite Image

 

第12页-17.PNG 


Result saving

7. Export Page

第12页-15.PNG 

                                                 Export page


① Storage location: Select the location of the image files automatically saved by the system, and the contents of the files will be displayed in a list format in the system.
② Open file: Open the selected directory.
③ Export: Select the export content and format to export the images.
④ Delete: Select the image file to be deleted, and it will be deleted.
⑤ Open: Select the corresponding file in the list to open and jump to the result page. You can operate on the image on the result page, or double-click the corresponding file to open it.
Return: Click "Return" to return to the main page.




Servicebio의 모든 제품을 만나 보세요!


Pathological Reagents | Consumables

Cell Biology Reagents | Consumables

Protein Biology Reagents | Consumables

Molecular Biology Reagents | Consumables

Antibodies

Biochemistry & Immunology

Recombinant Protein & Enzyme

Biochemical Reagents

Common Consumables

Instruments


Servicebio - Official Distributor in South Korea "Morebio" 한국 공식 대리점 "모아바이오"